Curve Tracer with
CRT |
|
|
Frequency
- Measurement Ranges
- 3 Voltage Ranges: Collector/Drain
Voltages £ 2V, 10V, 40V
±5%
- 3 Current ranges: Collector/Drain
Currents £ 2mA, 20mA,
200mA ±5%
- 3 Power Ranges: Output Power
£ 0.04W, 0.4W, 4W
±10%
- Base-/Gate-Voltages and Currents:
- Ib: 1µA to 10mA
- Vb: to 2V ±5%
- Vg: to 10V ±5%
|
Accuracy
Accuracy of Static
Values:
- Vc/d: ± (2% o.v.*) + 3
Dig.)
- Ic/d: ± (2% o.v. + 3 Dig.)
- Ib: ± (2% o.v. + 3 Dig.)
- Vb: ± (2% o.v. + 3 Dig.)
- Vg: ± (3% o.v. + 3 Dig.)
- b
to 1000: ± (5 % o.v. + 3 Dig.)
- b
to 100000: ± [(6 + 0.001 x b) % o.v. + 3 Dig.]
Accuracy of
Dynamic Values:
- h11: £ 1000 Ohm ± (12%
o.v.+ 3 Dig.)
- ³ 1000 Ohm ± [(12 + 0.001 meas. value) % o.v.+ 3 Dig.]
- h21: £ 1000 ± (12% o.v. + 3 Dig.)
- ³ 1000 ± [(12 + 0.001 meas.
value) % o.v.+ 3 Dig.]
- y21: £ 1000mS ± (12% o.v.. + 3 Dig.)
- h/y22: £ 1000mS ± (12% o.v.. + 3 Dig.)
|
Miscellanous
- Reference measurement values can be stored for component
selection.
- Cursor Measurements:
- Single mode: The Cursor marks the
position from which the measurement value is calculated.
- Tracking mode: Two Cursors mark
thepositions, from which the h/y-Parameter measurement values are
calculated.
- Evaluation of curves from: Diodes,
Zener Diodes, NPN/PNP-Transistors, FET/MOS-FET (N/P-Channel), Thyristors
- Display: 2x16-Digit, LCD
- Presentation of measurement values from a set of 5 curves on
CRT.
|
General Information
- CRT: D14-364GY/123 or
ER151-GH/-,
- 6" rectangular screen (8x10cm) internal graticule
- Acceleration voltage: approx.
2000V
- Trace rotation: adjustable on
front panel
- Line voltage: 100-240V AC ±10%, 50/60Hz
- Power consumption: approx. 36 Watt
at 50Hz.
- Min./Max. ambient temperature: 0°C...+40°C
- Protective system: Safety class I
(IEC 1010-1)
- Weight: approx. 5.6kg, color: techno-brown
- Cabinet: W 285, H 125, D 380 mm
- Lockable tilt handle
- *) o.v. = of value
| | |
| | |